Journal article icon

Journal article

Optimization and assessment of quantitative 124I imaging on a Philips Gemini dual GS PET/CT system.

Abstract:

PURPOSE: Quantitative (124)I PET imaging is challenging as (124)I has a complex decay scheme. In this study the performance of a Philips Gemini dual GS PET/CT system was optimized and assessed for (124)I. METHODS: The energy window giving the maximum noise equivalent count rate (NECR) and NEMA 2001-NU2 image quality were measured. The activity concentration (AC) accuracy of images calibrated using factors from (18)F and (124)I decaying source measurements were investigated. RESULTS: The energ...

Expand abstract

Actions


Access Document


Publisher copy:
10.1007/s00259-009-1099-8

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Oncology
Role:
Author
Journal:
European journal of nuclear medicine and molecular imaging
Volume:
36
Issue:
7
Pages:
1037-1048
Publication date:
2009-07-05
DOI:
EISSN:
1619-7089
ISSN:
1619-7070