Journal article icon

Journal article

Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years.

Abstract:

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1302/0301-620x.93b4.25605

Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author
Journal:
Journal of bone and joint surgery. British volume More from this journal
Volume:
93
Issue:
4
Pages:
470-475
Publication date:
2011-04-01
DOI:
ISSN:
0301-620X
Language:
English
Keywords:
Pubs id:
pubs:131299
UUID:
uuid:eb2970b3-2ad0-4010-969a-d25884c9284a
Local pid:
pubs:131299
Source identifiers:
131299
Deposit date:
2012-12-19

Terms of use


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record