Conference item icon

Conference item

Accuracy of a fluoroscopy technique for assessing patellar tracking

Abstract:
Accuracy of a novel fluoroscopy-based method of assessing patellar tracking was determined by comparing the pattern of patellar motion with respect to orientation (flexion, internal rotation, and lateral tilt) and translation (lateral, proximal, and anterior) with the pattern of patellar motion measured using Roentgen Stereophotogrammetric Analysis (RSA) in one cadaver knee specimen. Each pose in the patellar motion could be obtained from a single as well as multiple calibrated fluoroscopic images. The mean error (SD) varies from 0.73 (0.44) to 1.60 (0.48) degrees for patellar orientation and from 0.48 (0.37) to 1.20 (0.57) mm for patellar translation. These errors appear to be sufficiently low to identify clinically significant differences in patellar kinematics. © Springer-Verlag Berlin Heidelberg 2003.
Publication status:
Published

Actions


Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author


Host title:
MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2003, PT 1
Volume:
2878
Pages:
319-326
Publication date:
2003-01-01
EISSN:
1611-3349
ISSN:
0302-9743
ISBN:
3540204628


Pubs id:
pubs:108023
UUID:
uuid:e67c1cc3-3a40-4410-b58d-7535bd808d84
Local pid:
pubs:108023
Source identifiers:
108023
Deposit date:
2012-12-19

Terms of use



Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP