Optimisation of dark adaptation time required for mesopic microperimetry

Abstract:

Background Macular Integrity Assessment (MAIA) microperimetry is increasingly used in clinical and research settings to assess point retinal sensitivity and fixation stability. Testing occurs under mesopic conditions, commonly after a period of dark adaptation. Our aim was to identify the minimum length of adaptation required to optimise microperimetry performance.

Methods MAIA microperimetry using the 10-2 grid was performed on 40 right eyes of 40 healthy participants aged 18–73 with no ocular pathology and vision of at least 0.1 logMAR after ambient light exposure, with 0, 5, 10, 15, 20 and 30 min of adaptation in mesopic settings. Ten right eyes of 10 participants with choroideremia were also tested following 0 and 20 min of adaptation. We further tested 10 right eyes of 10 healthy participants after bright light exposure, with 0, 10 and 20 min of adaptation. We compared changes in threshold sensitivity and fixation stability across time points.

Results Microperimetry performance did not improve with increasing adaptation time in healthy participants or patients with choroideremia after ambient light exposure. After bright light exposure, we found microperimetry thresholds improved after 10 min of adaptation, but did not improve further at 20 min.

Conclusion Mesopic adaptation is not required before MAIA microperimetry after exposure to ambient light. Ten minutes of adaptation is sufficient after exposure to a bright light stimulus, such as ophthalmoscopy or retinal imaging. The brief time of dark adaptation required corresponds to cone adaptation curves and provides further evidence for cone-mediated central retinal function under mesopic conditions.

Publication status:
Published
Peer review status:
Peer reviewed

Actions

Access Document

Files:
Publisher copy:
10.1136/bjophthalmol-2018-312253

Authors

More by this author
Institution:
University of Oxford
Division:
MSD
Department:
Clinical Neurosciences
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences
Department:
Medical School Offices
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences
Department:
Clinical Neurosciences
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences
Department:
Clinical Neurosciences
Role:
Author


Publisher:
BMJ Publishing Group
Journal:
British Journal of Ophthalmology More from this journal
Volume:
103
Issue:
8
Pages:
1092-1098
Publication date:
2018-09-29
Acceptance date:
2018-08-28
DOI:
EISSN:
1468-2079
ISSN:
0007-1161


Keywords:
Pubs id:
pubs:911447
UUID:
uuid:de118b61-e05d-48d7-98b5-e5806ac74a98
Local pid:
pubs:911447
Source identifiers:
911447
Deposit date:
2018-08-31
ARK identifier:

Terms of use


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP