Conference item icon

Conference item

Accurate assessment of patellar tracking using fiducial and intensity-based fluoroscopic techniques.

Abstract:
Accuracies of a point-based and an intensity-based fluoroscopic methods of assessing patella tracking were determined by comparing the pattern of patellar motion with respect to orientation (flexion, internal rotation, and lateral tilt) and translation (lateral, proximal, and anterior) with the pattern of patellar motion measured using Roentgen stereophotogrammetric analysis in three cadaver knee specimens. Each pose in the patellar motion could be obtained from single as well as multiple calibrated fluoroscopic images. The errors using the intensity-based method were slightly higher than those of the point-based method, but they appear to be sufficiently low to detect clinically significant differences in patellar kinematics.
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1016/j.media.2004.06.011

Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author


Host title:
Medical image analysis
Volume:
8
Issue:
3
Pages:
343-351
Publication date:
2004-09-01
Event location:
England
DOI:
EISSN:
1361-8423
ISSN:
1361-8415