Journal article icon

Journal article

Characterization of an external cavity diode laser based ring cavity NICE-OHMS system.

Abstract:
The performance of an external cavity diode laser based noise immune cavity enhanced optical heterodyne molecular spectrometer is presented. To reduce the noise on the signal a ring cavity and a circuit to remove residual amplitude modulation on the pre-cavity laser radiation was implemented. We demonstrate a sensitivity of 4 x 10(-11) cm(-1) Hz(-1/2) using a cavity with a finesse of 2600 on a Doppler-broadened transition of CH(4) at 6610.063 cm(-1).
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1364/oe.17.009834

Authors


More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Chemistry
Sub department:
Physical & Theoretical Chem
Role:
Author
More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Chemistry
Sub department:
Physical & Theoretical Chem
Role:
Author


Publisher:
Optical Society of American (OSA)
Journal:
Optics express More from this journal
Volume:
17
Issue:
12
Pages:
9834-9839
Publication date:
2009-06-01