Journal article icon

Journal article

Acceptance Tests of more than 10 000 Photomultiplier Tubes for the multi-PMT Digital Optical Modules of the IceCube Upgrade

Abstract:
More than 10000 photomultiplier tubes (PMTs) with a diameter of 80 mm will be installed in multi-PMT Digital Optical Modules (mDOMs) of the IceCube Upgrade. These have been tested and pre-calibrated at two sites. A throughput of more than 1000 PMTs per week with both sites was achieved with a modular design of the testing facilities and highly automated testing procedures. The testing facilities can easily be adapted to other PMTs, such that they can, e.g., be re-used for testing the PMTs for IceCube-Gen2. Single photoelectron response, high voltage dependence, time resolution, prepulse, late pulse, afterpulse probabilities, and dark rates were measured for each PMT. We describe the design of the testing facilities, the testing procedures, and the results of the acceptance tests.
Publication status:
Published
Peer review status:
Peer reviewed

Actions


Access Document


Files:
Publisher copy:
10.1088/1748-0221/19/07/p07038

Authors



Publisher:
IOP Publishing
Journal:
Journal of Instrumentation More from this journal
Volume:
19
Issue:
07
Article number:
P07038
Publication date:
2024-07-29
Acceptance date:
2024-07-10
DOI:
EISSN:
1748-0221


Language:
English
Keywords:
Source identifiers:
2145366
Deposit date:
2024-07-29

Terms of use



Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record