Journal article
Achieving robust labeling above the circle of Willis with vessel-encoded arterial spin labeling
- Abstract:
-
Purpose
To improve the robustness of noninvasive vessel-selective perfusion imaging and angiography using vessel-encoded arterial spin labeling (VEASL) when applied to complex vascular geometries, such as above the circle of Willis (CoW) in the brain.
MethodsOur proposed improved optimized encoding scheme (IOES) better accounts for vascular geometry and the VEASL encoding process, leading to more SNR-efficient encodings than previous approaches. Pseudo-continuous arterial spin labeling (PCASL) parameters were optimized for a thinner labeling region, allowing tortuous vessels to be more accurately treated as single points within the labeling plane. Our optimized approach was compared to the original OES method above the CoW in healthy volunteers, with preliminary application in two Moyamoya patients.
ResultsIn simulation, the IOES improved SNR efficiency by approximately 10% and used longer wavelength encodings that are less sensitive to subject motion. The effective labeling thickness was reduced using optimized PCASL parameters, which maintained high labeling efficiency. In healthy volunteers, these improvements allowed for the separation of at least nine arteries and their downstream tissues, with more accurate vessel decoding and closer alignment between the measured VEASL signal modulation and the encoding design. Vascular territories consistent with angiography were found in the Moyamoya patients.
ConclusionsCombining IOES with optimized PCASL parameters, the vessel-decoding efficacy in a region with complex vascular geometry above the CoW was improved. The automated encoding design process and scan times under 6 min make it feasible to observe flow patterns above the CoW in clinical settings, particularly for studies of collateral circulation.
- Publication status:
- Published
- Peer review status:
- Peer reviewed
Actions
Access Document
- Files:
-
-
(Preview, Version of record, pdf, 6.8MB, Terms of use)
-
- Publisher copy:
- 10.1002/mrm.30542
Authors
- Publisher:
- Wiley
- Journal:
- Magnetic Resonance in Medicine More from this journal
- Volume:
- 94
- Issue:
- 4
- Pages:
- 1415-1431
- Place of publication:
- United States
- Publication date:
- 2025-07-01
- Acceptance date:
- 2025-04-07
- DOI:
- EISSN:
-
1522-2594
- ISSN:
-
0740-3194
- Language:
-
English
- Keywords:
- Pubs id:
-
2134236
- Local pid:
-
pubs:2134236
- Deposit date:
-
2025-07-14
- ARK identifier:
Terms of use
- Copyright holder:
- Li et al
- Copyright date:
- 2025
- Rights statement:
- © 2025 The Author(s). Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
- Licence:
- CC Attribution (CC BY)
If you are the owner of this record, you can report an update to it here: Report update to this record