Journal article
Internal model control of inhaled anaesthesia
- Abstract:
- This paper describes the design of a control system for the gas composition of the inhaled breath during anaesthesia. A model of the gas flow through the anaesthetic system has been developed which is used as a disturbance predictor in the control system. To cope with the considerable delays in the system, Internal Model Control is used and found to be effective. Two gases are controlled simultaneously with a time constant of 400ms, with strong disturbance rejection. © 2007 IEEE.
- Publication status:
- Published
Actions
Authors
- Journal:
- 2007 AMERICAN CONTROL CONFERENCE, VOLS 1-13 More from this journal
- Pages:
- 4958-4963
- Publication date:
- 2007-01-01
- DOI:
- ISSN:
-
0743-1619
- Language:
-
English
- Pubs id:
-
pubs:243293
- UUID:
-
uuid:84a0ea36-a5f9-447e-aeaf-dcc5f71ecf73
- Local pid:
-
pubs:243293
- Source identifiers:
-
243293
- Deposit date:
-
2012-12-19
Terms of use
- Copyright date:
- 2007
If you are the owner of this record, you can report an update to it here: Report update to this record