Journal article
OxyJet CPAP: an electricity-free low-cost emergency respiratory support device
- Abstract:
-
Background
Respiratory support devices in resource-limited settings should be inexpensive, portable, effective, and easy to use. Non-invasive respiratory support devices can reduce expensive ICU admissions, but these facilities are severely lacking in low-resource settings. Here, we describe the design and validation of a low-cost, portable, electricity-free, and 3D-printed continuous positive airway pressure (CPAP) device named 'OxyJet' that can provide non-invasive respiratory support outside the ICU. The OxyJet is uniquely built using off-the-shelf components and 3D printing technology, making it both inexpensive and easy to produce. OxyJet costs less than 10% of the price of a similar CPAP system. Inspired by the fundamental mechanics of gas ejectors and harnessing the potential energy of a high-pressure oxygen jet, the device can deliver a high flow of oxygenated air up to 65 L/min (approx.), a positive end-expiratory pressure (PEEP) within 5-15 cmH2O, and a fraction of inspired oxygen (FiO2) of up to 100%.Results
The device was bench-tested following UK-MHRA RMCPAP guidelines and tested on healthy volunteers (n = 5) and hypoxemic patients (n = 5). A comparative pilot study involving 23 hypoxemic adult patients conducted in Dhaka, Bangladesh, showed a significant improvement (p < 0.05) in the peripheral oxygen saturation (SpO2) of patients following the administration of OxyJet CPAP. The mean SpO2 increase was 12.0% (95% CI 10.8-13.2) with OxyJet versus 11.5% (95% CI 9.3-13.8) for standard CPAP (p = 0.695). The findings indicate the device's feasibility and short-term physiological effects comparable to those of standard CPAP systems. Further studies are required to confirm its clinical efficacy and broader utility in resource-limited settings.Conclusion
Our findings suggest that OxyJet CPAP has the potential to serve as an emergency respiratory support device outside the ICU, strengthening health systems in resource-limited settings.Methods
OxyJet's performance was first assessed through benchtop testing following the UK-MHRA RMCPAP protocol. This was followed by preliminary human testing in healthy volunteers and hypoxemic patients to evaluate safety and usability. A pilot feasibility study involving hypoxemic adult patients in Dhaka, Bangladesh, was then conducted to compare the device's physiological effects with those of standard CPAP therapy.
- Publication status:
- Published
- Peer review status:
- Peer reviewed
Actions
Access Document
- Files:
-
-
(Preview, Version of record, pdf, 1.4MB, Terms of use)
-
- Publisher copy:
- 10.1186/s12938-026-01524-7
Authors
- Publisher:
- BioMed Central
- Journal:
- BioMedical Engineering OnLine More from this journal
- Volume:
- 25
- Issue:
- 1
- Pages:
- 56
- Publication date:
- 2026-03-04
- DOI:
- EISSN:
-
1475-925X
- ISSN:
-
1475-925X
- Language:
-
English
- Keywords:
- Pubs id:
-
2427431
- Local pid:
-
pubs:2427431
- Source identifiers:
-
W7133530571
- Deposit date:
-
2026-06-01
- ARK identifier:
This ORA record was generated from metadata provided by an external service. It has not been edited by the ORA Team.