Journal article icon

Journal article

OxyJet CPAP: an electricity-free low-cost emergency respiratory support device

Abstract:

Background

Respiratory support devices in resource-limited settings should be inexpensive, portable, effective, and easy to use. Non-invasive respiratory support devices can reduce expensive ICU admissions, but these facilities are severely lacking in low-resource settings. Here, we describe the design and validation of a low-cost, portable, electricity-free, and 3D-printed continuous positive airway pressure (CPAP) device named 'OxyJet' that can provide non-invasive respiratory support outside the ICU. The OxyJet is uniquely built using off-the-shelf components and 3D printing technology, making it both inexpensive and easy to produce. OxyJet costs less than 10% of the price of a similar CPAP system. Inspired by the fundamental mechanics of gas ejectors and harnessing the potential energy of a high-pressure oxygen jet, the device can deliver a high flow of oxygenated air up to 65 L/min (approx.), a positive end-expiratory pressure (PEEP) within 5-15 cmH2O, and a fraction of inspired oxygen (FiO2) of up to 100%.

Results

The device was bench-tested following UK-MHRA RMCPAP guidelines and tested on healthy volunteers (n = 5) and hypoxemic patients (n = 5). A comparative pilot study involving 23 hypoxemic adult patients conducted in Dhaka, Bangladesh, showed a significant improvement (p < 0.05) in the peripheral oxygen saturation (SpO2) of patients following the administration of OxyJet CPAP. The mean SpO2 increase was 12.0% (95% CI 10.8-13.2) with OxyJet versus 11.5% (95% CI 9.3-13.8) for standard CPAP (p = 0.695). The findings indicate the device's feasibility and short-term physiological effects comparable to those of standard CPAP systems. Further studies are required to confirm its clinical efficacy and broader utility in resource-limited settings.

Conclusion

Our findings suggest that OxyJet CPAP has the potential to serve as an emergency respiratory support device outside the ICU, strengthening health systems in resource-limited settings.

Methods

OxyJet's performance was first assessed through benchtop testing following the UK-MHRA RMCPAP protocol. This was followed by preliminary human testing in healthy volunteers and hypoxemic patients to evaluate safety and usability. A pilot feasibility study involving hypoxemic adult patients in Dhaka, Bangladesh, was then conducted to compare the device's physiological effects with those of standard CPAP therapy.
Publication status:
Published
Peer review status:
Peer reviewed

Actions

Access Document

Files:
Publisher copy:
10.1186/s12938-026-01524-7

Authors


Publisher:
BioMed Central
Journal:
BioMedical Engineering OnLine More from this journal
Volume:
25
Issue:
1
Pages:
56
Publication date:
2026-03-04
DOI:
EISSN:
1475-925X
ISSN:
1475-925X


Language:
English
Keywords:
Pubs id:
2427431
Local pid:
pubs:2427431
Source identifiers:
W7133530571
Deposit date:
2026-06-01
ARK identifier:
This ORA record was generated from metadata provided by an external service. It has not been edited by the ORA Team.

Terms of use


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP