Journal article icon

Journal article

Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis.

Abstract:
A prospective double-blind, randomized, and controlled trial was conducted using roentgen stereophotogrammetric analysis; 54 total hip arthroplasty patients were randomized to receive either highly cross-linked polyethylene (HXLPE) or standard ultra-high-molecular-weight polyethylene (UHMWPE) liners. The 3-dimensional penetration of the liner was determined over 2 years. For the first 3 months, both polyethylene types had a rapid penetration rate (HXLPE: 0.22 mm, SD = 0.17 mm; UHMWPE: 0.21 mm, SD = 0.15 mm; P = .78). After 3 months, the HXLPE penetration rate (0.06 mm/y, SD = 0.06 mm/y) was significantly lower than the UHMWPE penetration rate (0.10 mm/y, SD = 0.07 mm/y; P = .04). The penetration in the first 3 months was probably caused by creep or bedding in; from 3 months onward, much of the penetration was probably caused by wear. We conclude that HXLPE has a 40% lower wear rate as compared with UHMWPE, suggesting that it will perform better in the long term.
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1016/j.arth.2006.12.117

Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author
More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author


Journal:
Journal of arthroplasty More from this journal
Volume:
23
Issue:
3
Pages:
337-343
Publication date:
2008-04-01
DOI:
EISSN:
1532-8406
ISSN:
0883-5403


Language:
English
Keywords:
Pubs id:
pubs:62556
UUID:
uuid:819f6aad-c5d7-462b-80d2-ae18c03617ad
Local pid:
pubs:62556
Source identifiers:
62556
Deposit date:
2012-12-19

Terms of use



Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record