Journal article icon

Journal article

Focused ultrasound hyperthermia for targeted drug release from thermosensitive liposomes: results from a phase I trial

Abstract:
Purpose To demonstrate the feasibility and safety of using focused ultrasound planning models to determine the treatment parameters needed to deliver volumetric mild hyperthermia for targeted drug delivery without real-time thermometry. Materials and Methods This study was part of the Targeted Doxorubicin, or TARDOX, phase I prospective trial of focused ultrasound–mediated, hyperthermia-triggered drug delivery to solid liver tumors (ClinicalTrials.gov identifier NCT02181075). Ten participants (age range, 49–68 years; average age, 60 years; four women) were treated from March 2015 to March 2017 by using a clinically approved focused ultrasound system to release doxorubicin from lyso-thermosensitive liposomes. Ultrasonic heating of target tumors (treated volume: 11–73 cm3 [mean ± standard deviation, 50 cm3 ± 26]) was monitored in six participants by using a minimally invasive temperature sensor; four participants were treated without real-time thermometry. For all participants, CT images were used with a patient-specific hyperthermia model to define focused ultrasound treatment plans. Feasibility was assessed by comparing model-prescribed focused ultrasound powers to those implemented for treatment. Safety was assessed by evaluating MR images and biopsy specimens for evidence of thermal ablation and monitoring adverse events. Results The mean difference between predicted and implemented treatment powers was −0.1 W ± 17.7 (n = 10). No evidence of focused ultrasound–related adverse effects, including thermal ablation, was found. Conclusion In this 10-participant study, the authors confirmed the feasibility of using focused ultrasound–mediated hyperthermia planning models to define treatment parameters that safely enabled targeted, noninvasive drug delivery to liver tumors while monitored with B-mode guidance and without real-time thermometry.
Publication status:
Published
Peer review status:
Peer reviewed

Actions


Access Document


Files:
Publisher copy:
10.1148/radiol.2018181445

Authors


More by this author
Institution:
University of Oxford
Division:
MPLS Division
Department:
Engineering Science
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences Division
Department:
Surgical Sciences
Role:
Author
More by this author
Institution:
University of Oxford
Division:
MPLS Division
Department:
Engineering Science
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences Division
Department:
Oncology
Role:
Author
More by this author
Institution:
University of Oxford
Division:
Medical Sciences Division
Department:
Oncology
Role:
Author


Publisher:
Radiological Society of North America
Journal:
Radiology More from this journal
Volume:
291
Issue:
1
Pages:
232–238
Publication date:
2019-01-15
Acceptance date:
2018-11-07
DOI:
EISSN:
1527-1315
ISSN:
0033-8419


Pubs id:
pubs:942835
UUID:
uuid:55cf1891-9d96-464b-9490-44cddbfb2e38
Local pid:
pubs:942835
Source identifiers:
942835
Deposit date:
2018-11-16

Terms of use



Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP