Journal article
Automated solid-phase extraction for determination of amodiaquine, chloroquine and metabolites in capillary blood on sampling paper by liquid chromatography
- Abstract:
- A bioanalytical method using solid-phase extraction and high-performance liquid chromatography (LC) has been developed and validated for the determination of amodiaquine, chloroquine and their metabolites in 100 μL blood applied to sampling paper. The intra-assay precision for all analytes was < 5% at 2000 nM, < 7% at 750 nM and < 10% at 100 nM. The inter-assay precision for all analytes was < 4% at 2000 nM, < 7% at 750 nM and < 12% at 100 nM. The lower limit of quantitation was 100 nM for all analytes. The limit of detection was 30 nM in 100 μL venous blood opplied to sampling paper.
- Publication status:
- Published
Actions
Authors
- Journal:
- CHROMATOGRAPHIA More from this journal
- Volume:
- 55
- Issue:
- 1-2
- Pages:
- 5-12
- Publication date:
- 2002-01-01
- DOI:
- EISSN:
-
1612-1112
- ISSN:
-
0009-5893
- Language:
-
English
- Keywords:
- Pubs id:
-
pubs:21103
- UUID:
-
uuid:46dbea10-5414-4288-b204-06a0844fd9e4
- Local pid:
-
pubs:21103
- Source identifiers:
-
21103
- Deposit date:
-
2012-12-19
Terms of use
- Copyright date:
- 2002
If you are the owner of this record, you can report an update to it here: Report update to this record