Journal article
Technical evaluation and usability of a quantitative G6PD POC test in cord blood: a mixed-methods study in a low-resource setting
- Abstract:
-
Objectives New point-of-care (POC) quantitative G6PD testing devices developed to provide safe radical cure for Plasmodium vivax malaria may be used to diagnose G6PD deficiency in newborns at risk of severe neonatal hyperbilirubinaemia, improving clinical care, and preventing related morbidity and mortality.
Methods We conducted a mixed-methods study analysing technical performance and usability of the ‘STANDARD G6PD’ Biosensor when used by trained midwives on cord blood samples at two rural clinics on the Thailand–Myanmar border.
Results In 307 cord blood samples, the Biosensor had a sensitivity of 1.000 (95% CI: 0.859 to 1.000) and a specificity of 0.993 (95% CI: 0.971 to 0.999) as compared with gold-standard spectrophotometry to diagnose G6PD-deficient newborns using a receiver operating characteristic (ROC) analysis-derived threshold of ≤4.8 IU/gHb. The Biosensor had a sensitivity of 0.727 (95% CI: 0.498 to 0.893) and specificity of 0.933 (95% CI: 0.876 to 0.969) for 30%–70% activity range in girls using ROC analysis-derived range of 4.9–9.9 IU/gHb. These thresholds allowed identification of all G6PD-deficient neonates and 80% of female neonates with intermediate phenotypes.
Need of phototherapy treatment for neonatal hyperbilirubinaemia was higher in neonates with deficient and intermediate phenotypes as diagnosed by either reference spectrophotometry or Biosensor.
Focus group discussions found high levels of learnability, willingness, satisfaction and suitability for the Biosensor in this setting. The staff valued the capacity of the Biosensor to identify newborns with G6PD deficiency early (‘We can know that early, we can counsel the parents about the chances of their children getting jaundice’) and at the POC, including in more rural settings (‘Because we can know the right result of the G6PD deficiency in a short time, especially for the clinic which does not have a lab’).
Conclusions The Biosensor is a suitable tool in this resource-constrained setting to identify newborns with abnormal G6PD phenotypes at increased risk of neonatal hyperbilirubinaemia.
- Publication status:
- Published
- Peer review status:
- Peer reviewed
Actions
Access Document
- Files:
-
-
(Preview, Version of record, pdf, 885.8KB, Terms of use)
-
- Publisher copy:
- 10.1136/bmjopen-2022-066529
Authors
- Publisher:
- BMJ Publishing Group
- Journal:
- BMJ Open More from this journal
- Volume:
- 12
- Issue:
- 12
- Article number:
- e066529
- Publication date:
- 2022-12-13
- Acceptance date:
- 2022-11-24
- DOI:
- EISSN:
-
2044-6055
- Pmid:
-
36523222
- Language:
-
English
- Keywords:
- Pubs id:
-
1315978
- Local pid:
-
pubs:1315978
- Deposit date:
-
2023-01-09
Terms of use
- Copyright holder:
- Bancone et al.
- Copyright date:
- 2022
- Rights statement:
- © Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY. Published by BMJ. This is an open access article distributed in accordance with the Creative Commons Attribution 4.0 Unported (CC BY 4.0) license, which permits others to copy, redistribute, remix, transform and build upon this work for any purpose, provided the original work is properly cited, a link to the licence is given, and indication of whether changes were made.
- Licence:
- CC Attribution (CC BY)
If you are the owner of this record, you can report an update to it here: Report update to this record