- Abstract:
-
Introduction Screening for G6PD deficiency can inform disease management including malaria. Treatment with the antimalarial drugs primaquine and tafenoquine can be guided by point-of-care testing for G6PD deficiency.
Methods and Findings
Data from similar clinical studies evaluating the performance of the STANDARD ™ G6PD Test (SD Biosensor, South Korea) conducted in Bangladesh, Brazil, Ethiopia, India, Thailand, the United Kingdom, and the United States were pooled. Test performance was assessed in a retrospective analysis on capillary and venous specimens. All study sites used spectrophotometry for reference G6PD testing, and either the HemoCue or complete blood count for reference hemoglobin measurement.
The sensitivity of the STANDARD ™ G6PD Test using the manufacturer thresholds for G6PD deficient and intermediate cases in capillary specimens from 4212 study participants was 100% (95% Confidence Interval (CI): 97.5%–100%) for G6PD deficient cases with <30% activity and 77% (95% CI 66.8%– 85.4%) for females with intermediate activity between 30%–70%. Specificity was 98.1% (95% CI 97.6%–98.5%) and 92.8% (95% CI 91.6%–93.9%) for G6PD deficient individuals and intermediate females, respectively. The majority (12/20) of G6PD intermediate females with false normal results had activity levels >60% on the reference assay. Negative predictive values for females with G6PD activity >60% was 99.6% (95% CI 99.1%–99.8%) on capillary specimens. Test sensitivity among 396 P. vivax malaria cases was 100% (69.2%–100.0%) for both deficient and intermediate cases. In the study population, a high proportion of those classified as G6PD deficient or intermediate resulted from true normal cases. Despite this, the majority cases would receive the correct medication and no true G6PD deficient cases would be treated inappropriately.
Conclusions
The STANDARD G6PD Test enables safe access to drugs which are contraindicated for individuals with G6PD deficiency. Operational considerations will inform test uptake in specific settings.
- Publication status:
- Published
- Peer review status:
- Not peer reviewed
Actions
Access Document
- Files:
-
-
(Preview, Pre-print, pdf, 650.5KB, Terms of use)
-
- Preprint server copy:
- 10.1101/2022.12.11.22282391
Authors
+ Wellcome Trust
More from this funder
- Funder identifier:
- https://ror.org/029chgv08
- Grant:
- 220211/A/20/A
- 200909/Z/16/Z
- 220211/Z/20/Z
- Preprint server:
- medRxiv
- Publication date:
- 2022-12-13
- DOI:
- Language:
-
English
- Pubs id:
-
1318453
- UUID:
-
uuid_3e549f4a-5dba-4a49-ad2b-d7185c44e825
- Local pid:
-
pubs:1318453
- Source identifiers:
-
W4311496078
- Deposit date:
-
2025-12-21
- ARK identifier:
Terms of use
- Copyright holder:
- Adissu et al
- Copyright date:
- 2022
- Rights statement:
- ©2022 The Author. The copyright holder for this preprint is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. It is made available under a CC-BY-NC 4.0 International license.
If you are the owner of this record, you can report an update to it here: Report update to this record