Conference item
Measurements of field enhancement introduced by a local electrode.
- Abstract:
-
Direct measurements of field enhancement introduced by a local electrode have been carried out on a scanning atom probe (SAP). The results show that an enhancement factor of more than 2 can be obtained simply by reducing the specimen-to-electrode distance from 1.5 mm to 10 microm. Further enhancement can be achieved if the electrode has a smaller aperture. This suggests that a specimen with tip radius of approximately 200 nm, which usually is too blunt to be analysed in a standard atom probe,...
Expand abstract
- Publication status:
- Published
Actions
Authors
Bibliographic Details
- Journal:
- Ultramicroscopy
- Volume:
- 89
- Issue:
- 1-3
- Pages:
- 163-167
- Publication date:
- 2001-10-01
- Event title:
- 46th International Field Emission Symposium