- Abstract:
-
Direct measurements of field enhancement introduced by a local electrode have been carried out on a scanning atom probe (SAP). The results show that an enhancement factor of more than 2 can be obtained simply by reducing the specimen-to-electrode distance from 1.5 mm to 10 microm. Further enhancement can be achieved if the electrode has a smaller aperture. This suggests that a specimen with tip radius of approximately 200 nm, which usually is too blunt to be analysed in a standard atom probe,...
Expand abstract - Publication status:
- Published
- Volume:
- 89
- Issue:
- 1-3
- Pages:
- 163-167
- Publication date:
- 2001-10-05
- DOI:
- EISSN:
-
1879-2723
- ISSN:
-
0304-3991
- URN:
-
uuid:3d9dd2d7-3f13-4f1a-b3d9-0bc1fa1d8f39
- Source identifiers:
-
23150
- Local pid:
- pubs:23150
- Keywords:
- Copyright date:
- 2001