Journal article
CryoEDM: A cryogenic experiment to measure the neutron electric dipole moment
- Abstract:
- We have constructed an instrument, CryoEDM, to measure the neutron electric dipole moment to a precision of 10-28 e cm at the Institut Laue-Langevin. The main characteristic is that it is operating entirely in a cryogenic environment, at temperatures of 0.7 K within superfluid helium. Ultracold neutrons are produced in a superthermal source and stored within the superfluid in a storage cell which is held in a magnetic and electric field. NMR measurements are carried out to look for any shifts in the neutron Larmor precession frequency associated with the electric field and the neutrons are detected in-situ in the superfluid. Low temperature SQUID magnetometry is used to monitor the magnetic field. We report on the current status of the project that is now being commissioned and give an outlook on the future exploitation of the instrument. © 2010 IOP Publishing Ltd.
Actions
Access Document
- Files:
-
-
(Preview, pdf, 769.2KB, Terms of use)
-
- Publisher copy:
- 10.1088/1742-6596/251/1/012055
Authors
- Journal:
- Journal of Physics: Conference Series More from this journal
- Volume:
- 251
- Issue:
- 1
- Publication date:
- 2010-01-01
- DOI:
- EISSN:
-
1742-6596
- ISSN:
-
1742-6588