Abstract:
PURPOSE: To study arterial spin labeling (ASL) MRI techniques and to investigate various problematic issues that still hinder the accurate and robust quantitative analysis of ASL data. MATERIALS AND METHODS: A pulsed-ASL (PASL) sequence was implemented on a 3-T imaging system and a protocol was developed for the measurement of perfusion based on fitting to a standard kinetic model. Both numerical simulations and multi-inversion time MRI data were analyzed. The effect of fitting a kinetic curve to a large region of interest (ROI) with a distribution of arterial transit times was compared to a pixel-by-pixel (PBP) method. RESULTS: It was found that a significant underestimation of perfusion of approximately 17+/-6% (P<0.001) occurs in gray matter, when comparing an ROI with a PBP analysis over a group of 12 healthy subjects. CONCLUSION: Analysis of ASL data based on a large ROI may suffer from inaccuracies arising from a distribution of transit times, implying that averaging of ASL kinetic data over such regions should therefore be avoided. When possible, a PBP fit should be performed.
Publication status:
Published

Actions

Access Document

Publisher copy:
10.1002/jmri.20329

Authors

More by this author
Institution:
University of Oxford
Division:
MSD
Department:
Clinical Neurosciences
Role:
Author


Journal:
Journal of magnetic resonance imaging : JMRI More from this journal
Volume:
21
Issue:
6
Pages:
676-682
Publication date:
2005-06-01
DOI:
EISSN:
1522-2586
ISSN:
1053-1807


Language:
English
Keywords:
Pubs id:
pubs:241968
UUID:
uuid:1445e260-30f9-454f-97e6-c43099329f10
Local pid:
pubs:241968
Source identifiers:
241968
Deposit date:
2012-12-19
ARK identifier:

Terms of use


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record

TO TOP