Journal article icon

Journal article

A low‐cost, microcontroller‐based gas delivery system for respiratory stimuli in MRI studies

Abstract:
Purpose

The design and validation of a low-cost, microcontroller-based gas delivery system that automates fixed inspired respiratory stimuli for MRI experiments.

Methods

The system uses three solenoid valves controlled by an Arduino-based circuit to switch between premixed medical gases according to predefined timings. By using the MRI scanner external timing signal, gas delivery can be synchronized with image acquisition. The system was constructed using commercial components, with a material cost ∼£650 GBP (∼$880 USD). The system was integrated with a single-use breathing circuit and evaluated using hypercapnic and hyperoxic stimuli. End-tidal oxygen (O2) and carbon dioxide (CO2) were measured using a respiratory gas analyzer and physiological responses were assessed using BOLD MRI at 3 T.

Results

The system delivered reliable, repeatable gas transitions during MRI-triggered protocols. During hypercapnia (elevated carbon dioxide level, 5% CO2) (n = 15), the mean increase in end-tidal CO2 was 8.7 ± 1.8 mmHg from a baseline of 32.2 ± 3.1 mmHg, producing a mean gray matter BOLD signal increase of 3.2% ± 1.7%. During hyperoxia (elevated oxygen level, 60.5% O2) (n = 15), the mean increase in end-tidal O2 was 292.3 ± 59.0 mmHg from a baseline of 114.5 ± 10.7 mmHg, with an associated BOLD signal change of 1.2% ± 1.7%. Across both protocols, respiratory and BOLD responses were consistent across participants.

Conclusion

This microcontroller-based system provides an inexpensive and reliable method for administering fixed inspired respiratory stimuli with automated MRI synchronization. It offers an intermediate option between manual systems and commercial gas blenders, making it well suited for technical and methodological studies in cerebrovascular reactivity, hyperoxia-BOLD, and related applications.

Publication status:
Published
Peer review status:
Peer reviewed

Actions

Access Document

Files:
Publisher copy:
10.1002/mrm.70543

Authors

More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Engineering Science
Sub department:
Institute of Biomedical Engineering
Oxford college:
St Hilda's College
Role:
Author
ORCID:
0000-0002-3034-8986
More by this author
Institution:
University of Oxford
Division:
MSD
Department:
Psychiatry
Role:
Author


More from this funder
Funder identifier:
https://ror.org/029chgv08
Grant:
203139/A/16/Z
Grant:
203139/Z/16/Z
More from this funder
Funder identifier:
10.13039/100000002
Grant:
203139/Z/16/Z
More from this funder
Funder identifier:
10.13039/501100020624
Grant:
NIHR203310


Publisher:
Wiley
Journal:
Magnetic Resonance in Medicine More from this journal
Volume:
96
Issue:
6
Pages:
2907-2913
Place of publication:
United States
Publication date:
2026-08-12
Acceptance date:
2026-07-20
DOI:
EISSN:
1522-2594
ISSN:
0740-3194
Pmid:
42590816


Language:
English
Pubs id:
2451940
Local pid:
pubs:2451940
Source identifiers:
W7202372206
Deposit date:
2026-11-02
ARK identifier:

Terms of use


Views and Downloads

Views and downloads will return soon






If you are the owner of this record, you can report an update to it here: Report update to this record