Journal article
Examination of the two-dimensional pupil function in coherent scanning microscopes using spherical particles.
- Abstract:
- The determination of the modulus of the pupil function in reflection acoustic and optical microscopy with the help of spherical particles is demonstrated. The theoretical examination shows that this method can give good results if the pupil function drops smoothly towards the edge. For a pupil with a sharp edge this technique can give the accurate pupil function only by imaging large spheres. The dependence of the accuracy of the method on the size of the spherical particle is analyzed. Numerical and experimental results obtained for the confocal scanning optical microscope and the scanning acoustic microscope are examined.
- Publication status:
- Published
Actions
Access Document
- Publisher copy:
- 10.1121/1.423268
Authors
- Publisher:
- American Institute of Physics Inc.
- Journal:
- Journal of the Acoustical Society of America More from this journal
- Volume:
- 104
- Issue:
- 1
- Pages:
- 181-191
- Publication date:
- 1998-07-01
- DOI:
- EISSN:
-
1520-8524
- ISSN:
-
0001-4966
- Language:
-
English
- Keywords:
- Pubs id:
-
pubs:62728
- UUID:
-
uuid:0d7d022f-2d85-4288-9f94-84fed999a38f
- Local pid:
-
pubs:62728
- Source identifiers:
-
62728
- Deposit date:
-
2012-12-19
- ARK identifier:
Terms of use
- Copyright date:
- 1998
If you are the owner of this record, you can report an update to it here: Report update to this record