Journal article
The accuracy of the Hewlett-Packard 47201A ear oximeter below 50% saturation.
- Abstract:
- The Hewlett-Packard 47201A ear oximeter has been shown to measure arterial oxygen saturation (SaO2) above 50% SaO2 with an accuracy of +/- 5%. Below 50% SaO2, the oximeter underestimates arterial saturation in a predictable way, thus allowing a correction factor to be used: true SaO2 = (oximeter reading + 50)/2.
- Publication status:
- Published
Actions
Authors
- Journal:
- Bulletin europeen de physiopathologie respiratoire More from this journal
- Volume:
- 18
- Issue:
- 5
- Pages:
- 791-794
- Publication date:
- 1982-01-01
- ISSN:
-
0395-3890
- Language:
-
English