Journal article
Calibration of gradient propagation delays for accurate two-dimensional radiofrequency pulses.
- Abstract:
-
Hardware-related delays between the requested and actual start times of the gradient waveforms on each physical axis are of particular importance for multidimensional selective excitation in which the synchronization of gradient and radiofrequency (RF) waveforms is critical. A method is proposed for the accurate calibration of gradient propagation delays to optimize the spatial accuracy of 2D RF pulses, although the results may also be used to reduce artifacts in other MR techniques. The sens...