Conference item
Research of superconducting tunnel junction X-ray detectors with direct signal observation method using a fast current readout system
- Abstract:
- We have developed a fast current readout system for STJ detectors with a fast current preamplifier by direct signal observation. The preamplifier has frequency bandwidth of 110 MHz with stable voltage bias operation. By the various experiments using a Ta STJ and Nb STJ, it is confirmed that the readout system is effective for the evaluation of STJs, research of the mechanism for signal production process and the X-ray detection with high counting rate. (C) 2000 Elsevier Science B.V. All rights reserved.
- Publication status:
- Published
Actions
Authors
Bibliographic Details
- Journal:
- NUCLEAR INSTRUMENTS and METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
- Volume:
- 444
- Issue:
- 1-2
- Pages:
- 124-128
- Publication date:
- 2000-04-11
- Event title:
- 8th International Workshop on Low-Temperature Detectors (LTD-8)
- DOI:
- ISSN:
-
0168-9002