Abstract:
We have developed a fast current readout system for STJ detectors with a fast current preamplifier by direct signal observation. The preamplifier has frequency bandwidth of 110 MHz with stable voltage bias operation. By the various experiments using a Ta STJ and Nb STJ, it is confirmed that the readout system is effective for the evaluation of STJs, research of the mechanism for signal production process and the X-ray detection with high counting rate. (C) 2000 Elsevier Science B.V. All rights reserved.