Journal article
Glycosylated haemoglobin: comparison of five different methods, including measurement on capillary blood samples.
- Abstract:
-
Glycosylated haemoglobin was measured in venous blood samples and in blood collected in 'Unistep' bottles by isoelectric focusing (IEF), as the reference method, and by electroendosmosis (EEO), the thiobarbituric acid method (TBA), ion-exchange chromatography (IEC) and affinity chromatography (AC). Isoelectric focusing, electroendosmosis and thiobarbituric acid gave similar results. Affinity chromatography gave lower results than isoelectric focusing for normal values but similar results for ...
Expand abstract
- Publication status:
- Published
Actions
Authors
Bibliographic Details
- Journal:
- Annals of clinical biochemistry
- Volume:
- 23 ( Pt 1)
- Issue:
- 1
- Pages:
- 85-91
- Publication date:
- 1986-01-01
- DOI:
- EISSN:
-
1758-1001
- ISSN:
-
0004-5632
Item Description
- Language:
- English
- Keywords:
- Pubs id:
-
pubs:12885
- UUID:
-
uuid:05dae233-8d7f-474e-ba2b-882e664f797e
- Local pid:
- pubs:12885
- Source identifiers:
-
12885
- Deposit date:
- 2012-12-19
Terms of use
- Copyright date:
- 1986
Metrics
If you are the owner of this record, you can report an update to it here: Report update to this record