- Abstract:
- Nanoflow electrospray mass spectrometry has been applied previously to investigate noncovalent protein-protein and protein-ligand interactions. Here we evaluate a commercial microchip device for this application. We show that the microchip can be used to obtain mass spectra of the noncovalent tetramer transthyretin. The device showed a 10-fold increase in signal stability compared with a nanoflow capillary and a high level of nozzle-to-nozzle reproducibility. Binding of the natural ligand thyroxine was clearly observed, and a range of small molecules proposed as inhibitors of transthyretin amyloidosis were shown to be effective in stabilizing the tetramer. We propose that measuring the ability of small molecules to stabilize protein complexes using this automated microchip technology will enable high-throughput screening of multi-protein complexes by mass spectrometry.
- Publication status:
- Published
Actions
Access Document
- Publisher copy:
- 10.1021/ac034251c
Authors
- Journal:
- Analytical chemistry More from this journal
- Volume:
- 75
- Issue:
- 18
- Pages:
- 4937-4941
- Publication date:
- 2003-09-01
- DOI:
- EISSN:
-
1520-6882
- ISSN:
-
0003-2700
- Language:
-
English
- Keywords:
- Pubs id:
-
pubs:59254
- UUID:
-
uuid:04bb3feb-5ba5-41ae-be08-b793f1102de9
- Local pid:
-
pubs:59254
- Source identifiers:
-
59254
- Deposit date:
-
2012-12-19
- ARK identifier:
Terms of use
- Copyright date:
- 2003
If you are the owner of this record, you can report an update to it here: Report update to this record