Journal article icon

Journal article

Effects of acetabular resurfacing component material and fixation on the strain distribution in the pelvis.

Abstract:

A 3D finite element (FE) model of an implanted pelvis was developed as part of a project investigating an all-polymer hip resurfacing design. The model was used to compare this novel design with a metal-on-metal design in current use and a metal-on-polymer design typical of early resurfacing implants. The model included forces representing the actions of 22 muscles as well as variable cancellous bone stiffness and variable cortical shell thickness. The hip joint reaction force was applied via...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1243/09544110260138727

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MPLS, Engineering Science
Northmore-Ball, MD More by this author
Tanner, KE More by this author
Journal:
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
Volume:
216
Issue:
4
Pages:
237-245
Publication date:
2002
DOI:
EISSN:
2041-3033
ISSN:
0954-4119
URN:
uuid:fd57f645-9042-48d3-9dfc-cf6f8c4ae486
Source identifiers:
428251
Local pid:
pubs:428251

Terms of use


Metrics



If you are the owner of this record, you can report an update to it here: Report update to this record