Journal article icon

Journal article

Roentgen stereophotogrammetric analysis for assessing migration of total hip replacement femoral components.

Abstract:

A new Roentgen stereophotogrammetric analysis system, using a biplane technique, has been developed to determine the migration and rotation of total hip replacement (THR) femoral components in three dimensions. Stainless steel marker balls were injected into the femur during the operation. The patients stood within a calibration frame during the X-ray. The two exposures were taken consecutively allowing radio-opaque shutters to be moved in front of the films to prevent fogging. Studies with a...

Expand abstract

Actions


Access Document


Authors


More by this author
Institution:
University of Oxford
Division:
MSD
Department:
NDORMS
Role:
Author
Journal:
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine More from this journal
Volume:
209
Issue:
3
Pages:
169-175
Publication date:
1995-01-01
DOI:
EISSN:
2041-3033
ISSN:
0954-4119
Language:
English
Keywords:
Pubs id:
pubs:62654
UUID:
uuid:f97cedec-af9b-46b5-9845-03ac87b991f4
Local pid:
pubs:62654
Source identifiers:
62654
Deposit date: