Experimental validation of a tractable numerical model for focused ultrasound heating in flow-through tissue phantoms

Abstract:

Heating from high intensity focused ultrasound (HIFU) can be used to control bleeding, both from individual blood vessels as well as from gross damage to the capillary bed. The presence of vascularity can limit one’s ability to elevate the temperature owing to convective heat transport. In an effort to better understand the heating process in tissues with vascular structure we have developed a numerical simulation that couples models for ultrasound propagation...

Expand abstract
Publication status:
Published
Peer review status:
Peer reviewed

Actions


Access Document


Publisher copy:
10.1121/1.1787124

Authors


More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Engineering Science
Oxford college:
Magdalen College
Role:
Author
More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Engineering Science
Oxford college:
Harris Manchester College
Role:
Author
ORCID:
0000-0001-6394-5667
Publisher:
Acoustical Society of America Publisher's website
Journal:
Journal of the Acoustical Society of America Journal website
Volume:
116
Issue:
4
Pages:
2451-2458
Publication date:
2004-10-06
Acceptance date:
2004-07-11
DOI:
EISSN:
1520-8524
ISSN:
0001-4966
Pubs id:
pubs:327647
URN:
uri:f8d5307b-86db-41d7-a8cb-0df0effb5a50
UUID:
uuid:f8d5307b-86db-41d7-a8cb-0df0effb5a50
Local pid:
pubs:327647
Language:
English