Journal article icon

Journal article

Human cardiac 31P magnetic resonance spectroscopy at 7 Tesla.

Abstract:

PURPOSE: Phosphorus magnetic resonance spectroscopy ((31)P-MRS) affords unique insight into cardiac energetics but has a low intrinsic signal-to-noise ratio (SNR) in humans. Theory predicts an increased (31)P-MRS SNR at 7T, offering exciting possibilities to better investigate cardiac metabolism. We therefore compare the performance of human cardiac (31)P-MRS at 7T to 3T, and measure T1s for (31)P metabolites at 7T. METHODS: Matched (31)P-MRS data were acquired at 3T and 7T, on nine normal vo...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1002/mrm.24922

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, RDM, RDM Cardiovascular Medicine
Clarke, WT More by this author
Vaughan, JT More by this author
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, RDM, RDM Strategic
Expand authors...
Journal:
Magnetic resonance in medicine
Volume:
72
Issue:
2
Pages:
304-315
Publication date:
2014-08-05
DOI:
EISSN:
1522-2594
ISSN:
0740-3194
URN:
uuid:f2894857-e7b6-46f6-b4f0-e422feb23291
Source identifiers:
420913
Local pid:
pubs:420913

Terms of use


Metrics



If you are the owner of this record, you can report an update to it here: Report update to this record