Relative effectiveness of clinic and home blood pressure monitoring compared with ambulatory blood pressure monitoring in diagnosis of hypertension: systematic review.

Abstract:

OBJECTIVE: To determine the relative accuracy of clinic measurements and home blood pressure monitoring compared with ambulatory blood pressure monitoring as a reference standard for the diagnosis of hypertension. DESIGN: Systematic review with meta-analysis with hierarchical summary receiver operating characteristic models. Methodological quality was appraised, including evidence of validation of blood pressure measurement equipment. DATA SOURCES: Medline (from 1966), Embase (from 1980), Coc...