Journal article
Traveling-wave ion mobility mass spectrometry of protein complexes: accurate calibrated collision cross-sections of human insulin oligomers.
- Abstract:
-
RATIONALE: The collision cross-section (Ω) of a protein or protein complex ion can be measured using traveling-wave (T-wave) ion mobility (IM) mass spectrometry (MS) via calibration with compounds of known Ω. The T-wave Ω-values depend strongly on instrument parameters and calibrant selection. Optimization of instrument parameters and calibration standards are crucial for obtaining accurate T-wave Ω-values. METHODS: Human insulin and the fast-acting insulin aspart under native-like condition...
Expand abstract
- Publication status:
- Published
Actions
Authors
Bibliographic Details
- Journal:
- Rapid communications in mass spectrometry : RCM More from this journal
- Volume:
- 26
- Issue:
- 10
- Pages:
- 1181-1193
- Publication date:
- 2012-05-01
- DOI:
- EISSN:
-
1097-0231
- ISSN:
-
0951-4198
Item Description
- Language:
-
English
- Keywords:
- Pubs id:
-
pubs:325391
- UUID:
-
uuid:de4f7e44-7fae-4d01-97f2-c1adde43b5c2
- Local pid:
-
pubs:325391
- Source identifiers:
-
325391
- Deposit date:
-
2013-11-16
Terms of use
- Copyright date:
- 2012
Metrics