Journal article
Analysis of the uncertainty in microbubble characterization
- Abstract:
-
There is increasing interest in the use of microbubble contrast agents for quantitative imaging applications such as perfusion and blood pressure measurement. The response of a microbubble to ultrasound excitation is, however, extremely sensitive to its size, the properties of its coating and the characteristics of the sound field and surrounding environment. Hence the results of microbubble characterization experiments can be significantly affected by experimental uncertainties, and this can...
Expand abstract
- Publication status:
- Published
- Peer review status:
- Peer reviewed
Actions
Access Document
- Files:
-
-
(Version of record, pdf, 681.6KB)
-
- Publisher copy:
- 10.1016/j.ultrasmedbio.2016.01.005
Authors
Funding
Bibliographic Details
- Publisher:
- Elsevier Publisher's website
- Journal:
- Ultrasound in Medicine and Biology Journal website
- Volume:
- 42
- Issue:
- 6
- Pages:
- 1412–1418
- Publication date:
- 2016-03-01
- Acceptance date:
- 2016-01-11
- DOI:
- EISSN:
-
1879-291X
- ISSN:
-
0301-5629
Item Description
- Language:
- English
- Keywords:
- Pubs id:
-
pubs:611732
- UUID:
-
uuid:d5b098d4-734d-4a89-a145-3459607c17e0
- Local pid:
- pubs:611732
- Source identifiers:
-
611732
- Deposit date:
- 2016-04-14
Terms of use
- Copyright holder:
- Stride et al
- Copyright date:
- 2016
- Notes:
- Copyright © 2016 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
- Licence:
- CC Attribution (CC BY)
Metrics
If you are the owner of this record, you can report an update to it here: Report update to this record