Correction of multi-collector-ICP-MS instrumental biases in high-precision uranium-thorium chronology

Abstract:

We present a method for the calibration of in-house thorium isotope standards for use in U-Th chronology, along with an assessment of instrumental biases on the Nu Plasma multi-collector inductively-coupled-plasma mass spectrometer. Uranium and thorium mass fractionation is found to be very strongly coupled, with no evidence of differences in mass bias behaviour. Both Th-isotope and U-Th fractionation are most closely approximated by a linear-law fractionation model. Any deviations from linea...