Phase contrast ultrashort TE: A more reliable technique for measurement of high-velocity turbulent stenotic jets.

Abstract:

Accurate measurement of peak velocity is critical to the assessment of patients with stenotic valvular disease. Conventional phase contrast (PC) methods for imaging high-velocity jets in aortic stenosis are susceptible to intravoxel dephasing signal loss, which can result in unreliable measurements. The most effective method for reducing intravoxel dephasing is to shorten the echo time (TE); however, the amount that TE can be shortened in conventional sequences is limited. A new sequence inco...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1002/mrm.22051

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, RDM, Cardiovascular Medicine
Role:
Author
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, RDM, Cardiovascular Medicine, BHF Centre of Research Excellence
Role:
Author
Journal:
Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine
Volume:
62
Issue:
3
Pages:
626-636
Publication date:
2009-09-05
DOI:
EISSN:
1522-2594
ISSN:
0740-3194
URN: