Detection of As(III) via oxidation to As(V) using platinum nanoparticle modified glassy carbon electrodes: arsenic detection without interference from copper.

Abstract:

The electrochemical detection of As(III) was investigated on a platinum nanoparticle modified glassy carbon electrode in 1 M aqueous HClO4. Platinum nanoparticle modified glassy carbon electrodes were prepared by potential cycling in 0.1 M aqueous KCl containing 1 mM K2PtCl6. In each potential cycle, the potential was held at + 0.5 V for 0.01 s and at -0.7 V for 10 s. 25 cycles were optimally used to prepare the electrodes. The resulting electrode surfaces were characterized with AFM. The res...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1039/b513686e

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MPLS, Chemistry, Physical and Theoretical Chem
Journal:
The Analyst
Volume:
131
Issue:
4
Pages:
516-521
Publication date:
2006-04-05
DOI:
EISSN:
1364-5528
ISSN:
0003-2654