- Abstract:
-
A dual-channel, curved-crystal spectrograph was designed to measure time-integrated x-ray spectra in the approximately 1.5 to 2 keV range (6.2-8.2 A wavelength) from small-mass, thin-foil targets irradiated by the VULCAN petawatt laser focused up to 4x10(20) W/cm(2). The spectrograph consists of two cylindrically curved potassium-acid-phthalate crystals bent in the meridional plane to increase the spectral range by a factor of approximately 10 compared to a flat crystal. The device acquires s...
Expand abstract - Publication status:
- Published
- Journal:
- The Review of scientific instruments
- Volume:
- 80
- Issue:
- 8
- Pages:
- 083501
- Publication date:
- 2009-08-05
- DOI:
- EISSN:
-
1089-7623
- ISSN:
-
0034-6748
- URN:
-
uuid:ac041257-e8bc-4517-bc5e-498187305960
- Source identifiers:
-
1573
- Local pid:
- pubs:1573
- Language:
- English
- Copyright date:
- 2009
Journal article
A dual-channel, curved-crystal spectrograph for petawatt laser, x-ray backlighter source studies.