Validity and reliability of short-term heart-rate variability from the Polar S810.

Abstract:

PURPOSE: : To assess the validity and the reliability of short-term resting heart-rate variability (HRV) measures obtained using the Polar S810 heart-rate monitor and accompanying software. METHODS: : Measures of HRV were obtained from 5-min R to R wave (RR) interval data for 19 males and 14 females during 10 min of quiet rest on three separate occasions at 1-wk intervals using the Polar S810. Criterion measures of HRV were obtained simultaneously using the CardioPerfect (CP; Medical Graphics...

Expand abstract
Publication status:
Published

Actions


Access Document


Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Primary Care Health Sciences, PHC-Madox
Role:
Author
Expand authors...
Journal:
Medicine and science in sports and exercise
Volume:
41
Issue:
1
Pages:
243-250
Publication date:
2009-01-05
DOI:
EISSN:
1530-0315
ISSN:
0195-9131