High throughput assay for the determination of lumefantrine in plasma.

Abstract:

A high throughput bioanalytical assay for the determination of lumefantrine in plasma has been developed and validated extensively. The within-day precisions for lumefantrine were 5.2, 3.5 and 2.5% at 200, 2000 and 15000 ng/mL, respectively. The between-day precisions were 4.0, 2.8 and 3.1% at 200, 2000 and 15000 ng/mL, respectively. The lower limits of quantification (LLOQ) and the limits of detection (LOD) were 25 and 10 ng/mL, respectively using 0.250 mL plasma. The average recovery of lum...

Expand abstract
Publication status:
Published

Actions


Access Document


Authors


Annerberg, A More by this author
Singtoroj, T More by this author
Tipmanee, P More by this author
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Clinical Medicine, Thailand/Laos MOP
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Clinical Medicine, Thailand/Laos MOP
Expand authors...
Journal:
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Volume:
822
Issue:
1-2
Pages:
330-333