The slip-and-slide algorithm: a refinement protocol for detector geometry

Abstract:

Geometry correction is traditionally plagued by mis-fitting of correlated parameters, leading to local minima which prevent further improvements. Segmented detectors pose an enhanced risk of mis-fitting: even a minor confusion of detector distance and panel separation can prevent improvement in data quality. The slip-and-slide algorithm breaks down effects of the correlated parameters and their associated target functions in a fundamental shift in the approach to the problem. Parameters are n...

Expand abstract
Publication status:
Published
Peer review status:
Peer reviewed
Version:
Publisher's version

Actions


Access Document


Files:
Publisher copy:
10.1107/S1600577517013327

Authors


More by this author
Institution:
University of Oxford
Division:
Medical Sciences Division
Department:
NDM
Subgroup:
Human Genetics Wt Centre
Oxford college:
Hertford College
ORCID:
0000-0002-3426-4210
Publisher:
International Union of Crystallography Publisher's website
Journal:
Journal of Synchrotron Radiation Journal website
Volume:
24
Issue:
Pt 6
Pages:
1152-1162