Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter.

Abstract:

BACKGROUND: A pretrial clinical improvement project for the BOOST-II UK trial of oxygen saturation targeting revealed an artefact affecting saturation profiles obtained from the Masimo Set Radical pulse oximeter. METHODS: Saturation was recorded every 10 s for up to 2 weeks in 176 oxygen dependent preterm infants in 35 UK and Irish neonatal units between August 2006 and April 2009 using Masimo SET Radical pulse oximeters. Frequency distributions of % time at each saturation were plotted. An a...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1136/adc.2010.206011

Authors


Expand authors...
Journal:
Archives of disease in childhood. Fetal and neonatal edition
Volume:
96
Issue:
6
Pages:
F429-F433