Journal article
The effect of insulator nano-sheath thickness on the steady state current at a micro-disc electrode
- Abstract:
-
© 2015 Elsevier B.V. All rights reserved. Abstract The relative size of the insulating sheath to electrode area at a micro-disc electrode can lead to significant perturbations in the steady state current observed. A minimum, constant steady state current value is realised once the sheath thickness is greater than twice the radius Δl>2rd. However, as the sheath thickness decreases below this value, the observed current increases. In this paper a theoretical model is presented, allowing for ...
Expand abstract
- Publication status:
- Published
- Peer review status:
- Peer reviewed