Journal article icon

Journal article

Sites of deposition of aqueous aerosols: a study of efficiency of delivery systems for lung ventilation imaging in man.

Abstract:

A study was made of the deposition of 99Tcm-DTPA aerosol in the components of a jet nebulizer-based aerosol production system. Three impaction devices were compared: a ball-bearing separator, a virtual impactor and a step separator. In addition a comparison was made of two types of tubing which carried aerosol from nebulizer to mouthpiece: corrugated and smooth-walled tubing. The retention of aerosol following inhalation was measured in five normal volunteers using different patterns of breat...

Expand abstract
Publication status:
Published

Actions


Access Document


Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, Clinical Medicine, CCMP
Role:
Author
Journal:
Nuclear medicine communications
Volume:
10
Issue:
3
Pages:
171-180
Publication date:
1989-03-05
DOI:
EISSN:
1473-5628
ISSN:
0143-3636
URN:
uuid:778664c9-5946-4f8e-936a-6411c63653bf
Source identifiers:
27263
Local pid:
pubs:27263

Terms of use


Metrics


Views and Downloads






If you are the owner of this record, you can report an update to it here: Report update to this record