Journal article icon

Journal article

Multi-sensor fusion for robust computation of breathing rate

Abstract:
Respiratory information can be obtained from the changes in electrical impedance across the chest, the electrocardiogram or the changes in light absorption across the finger. Running estimates of the breathing rate from each of these can be obtained from independent Kalman filters. A robust estimate of breathing rate is derived by fusing the outputs of these independent Kalman filters using a weighting calculated from the squared differences between each prediction from the filters and the corresponding measurements.
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1049/el:20020773

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MPLS, Engineering Science, Biomedical Research Centre
Townsend, N More by this author
Journal:
ELECTRONICS LETTERS
Volume:
38
Issue:
22
Pages:
1314-1316
Publication date:
2002-10-24
DOI:
ISSN:
0013-5194
URN:
uuid:74739ea8-0ea5-4060-8193-02813e5587d6
Source identifiers:
61586
Local pid:
pubs:61586
Language:
English

Terms of use


Metrics



If you are the owner of this record, you can report an update to it here: Report update to this record