Journal article
In vivo measurement of total knee replacement wear.
- Abstract:
-
Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control ...
Expand abstract
- Publication status:
- Published
Actions
Authors
Bibliographic Details
- Journal:
- Knee
- Volume:
- 11
- Issue:
- 3
- Pages:
- 183-187
- Publication date:
- 2004-06-01
- DOI:
- EISSN:
-
1873-5800
- ISSN:
-
0968-0160