Journal article icon

Journal article

Pressure-dependent attenuation with microbubbles at low mechanical index.

Abstract:

It has previously been shown that the attenuation of ultrasound (US) by microbubble contrast agents is dependent on acoustic pressure (Chen et al. 2002). Although previous studies have modelled the pressure-dependence of attenuation in single bubbles, this paper investigates this subject by considering a bulk volume of bubbles together with other linear attenuators. Specifically, a new pressure-dependent attenuation model for an inhomogeneous volume of attenuators is proposed. In this model, ...

Expand abstract
Publication status:
Published

Actions


Access Document


Authors


More by this author
Institution:
University of Oxford
Division:
MPLS
Department:
Engineering Science
Role:
Author
Journal:
Ultrasound in medicine and biology
Volume:
31
Issue:
3
Pages:
377-384
Publication date:
2005-03-01
DOI:
EISSN:
1879-291X
ISSN:
0301-5629