Journal article icon

Journal article

In vitro comparison of the effects of rough and polished stem surface finish on pressure generation in cemented hip arthroplasty.

Abstract:

BACKGROUND AND PURPOSE: High pressures around implants can cause bone lysis and loosening. We investigated how pressures are generated around cemented femoral stems. METHOD: We compared the pressures generated by rough and polished tapered stems at their cement interfaces, in an in vitro model, before and after 1 million load cycles. RESULTS: At the start of the study, the loading of both polished and rough stems generated interface pressures that were not statistically significantly differe...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.3109/17453670902876755

Authors


Bartlett, GE More by this author
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, NDORMS
More by this author
Institution:
University of Oxford
Department:
Oxford, MSD, NDORMS
Journal:
Acta orthopaedica
Volume:
80
Issue:
2
Pages:
144-149
Publication date:
2009-04-05
DOI:
EISSN:
1745-3682
ISSN:
1745-3674
URN:
uuid:585a4ee7-8336-4e31-81b2-e6963408080a
Source identifiers:
108253
Local pid:
pubs:108253

Terms of use