Multimode electrometer for atmospheric ion measurements

Abstract:

A computer-controlled electrometer capable of either high impedance voltage measurements or femtoampere current measurements has been developed. It is intended for use with a Gerdien ion counter in atmospheric temperatures. Temperature variations can cause significant thermal errors arising from offset voltage and bias current changes that compromise the measurements. Use of high impedance switching permits the offset voltage and input bias currents of the electrometer to be measured, and all...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1063/1.1327303

Authors


Harrison, RG More by this author
Journal:
REVIEW OF SCIENTIFIC INSTRUMENTS
Volume:
71
Issue:
12
Pages:
4683-4685