We report the electroanalytical determination of lead by anodic stripping voltammetry at in-situ-formed, bismuth-film-modified, boron-doped diamond electrodes. Detection limits in 0.1 mol L(-1) nitric acid solution of 9.6 x 10(-8) mol L(-1) (0.2 ppb) and 1.1 x 10(-8) mol L(-1) (2.3 ppb) were obtained after 60 and 300 s deposition times, respectively. An acoustically assisted deposition procedure was also investigated and found to result in improved limits of detection of 2.6 x 10(-8) mol L(-1...