Abstract:

As part of a feasibility study into the use of novel electron detectors for an X-ray photoelectron emission microscope (XPEEM), we have characterised a detector based on microchannel plates (MCPs), a phosphor screen and a CCD camera. For XPEEM, an imaging detector is required for electrons in the energy range 10-20 keV. This type of detector is a standard fitment on commercial instruments and we have studied its performance in some detail in order to provide a baseline against which to evalua...