Trace species detection in the near infrared using Fourier transform broadband cavity enhanced absorption spectroscopy: initial studies on potential breath analytes.

Abstract:

Cavity enhanced absorption measurements have been made of several species that absorb light between 1.5 and 1.7 µm using both a supercontinuum source and superluminescent light emitting diodes. A system based upon an optical enhancement cavity of relatively high finesse, consisting of mirrors of reflectivity ∼99.98%, and a Fourier transform spectrometer, is demonstrated. Spectra are recorded of isoprene, butadiene, acetone and methane, highlighting problems with spectral interference and unam...

Expand abstract
Publication status:
Published

Actions


Access Document


Publisher copy:
10.1039/c0an00462f

Authors


More by this author
Institution:
University of Oxford
Department:
Oxford, MPLS, Chemistry, Physical and Theoretical Chem
Role:
Author
More by this author
Institution:
University of Oxford
Department:
Oxford, MPLS, Chemistry, Physical and Theoretical Chem
Role:
Author
Expand authors...
Journal:
The Analyst
Volume:
136
Issue:
4
Pages:
801-806
Publication date:
2011-02-05
DOI:
EISSN:
1364-5528
ISSN:
0003-2654